Spirometry Resources Centre

Quality assurance, infection protection and control

Quality assurance, infection protection and control

Aims

  1. To understand the rationale for quality assurance including ability to correctly describe, document, and ensure quality standards and practice
  2. To understand strategies for the prevention of infection transmission

Overview

1. Provide strategies for quality assurance

  • Requirements for equipment quality control
  • Calibration and verification
  • Importance of recognising device measurement errors during calibration and quality checks

2. Provide strategies for hygiene and infection protection and control

  • Precautions to protect staff and patients
  • Rationale for regular cleaning
  • Describe methods for prevention of infection transmission
  • Performance and record of infection control procedures

Quality Assurance

  • Quality Control (QC) is the process of measuring and monitoring the quality of the data.
  • Quality Assurance (QA) is a group of routines and interventions to ensure the data is high quality.
  • Good QA systems use a robust series of checks before during and after the patients' visit to ensure the results will be accurate and as precise as possible.
  • QA and QC are valuable tools to ensure that those performing the test can resolve arising issues, providing accurate and meaningful results (best representation of the patient's clinical status).
  • A QA/QC programme can help identify and minimise sources of error.
Pre procedure notes Familiarise yourself with the equipment and manufacturer instructions on calibration or calibration verification.

Spirometer equipment ideally is ISO 26782 compliant and calibration syringe has an in-date certificate. The syringe is always checked for leaks (daily inspection and monthly leak check must be documented). Checks all parts are in good working order before this procedure. Note that room temperature and syringe temperature may negatively affect results. Refer to latest national and international standards. If using software consult the manufacturer instructions for use. Refer to latest national and international standards.
Calibration types A calibration verification is the procedure used to validate that the device is within calibration limits (i.e., =/-3% [accuracy tolerance, =/-2.5% for spirometers plus =/-0.5% for calibration syringes]). In calibration mode the equipment will exclude the BTPS factor to the results.
Note on pre-calibrated sensors Precalibrated sensors must undergo calibration verification. Manufacturers must specify the action to be taken if a pre-calibrated device fails the calibration verification.
Setup procedure Always use a filter during the calibration of flow sensing equipment. Connect the calibrator (3 L Syringe) securely to the spirometer sensor without causing allowing air to flow through the device. If variable flow is detected during the zero-flow setting procedure or if the zero level has changed significantly, the zero-flow setting procedure must be repeated.
Perform procedure Fill the syringe completely with air and attach the calibration syringe firmly to the equipment sensor (always use a filter during the calibration of flow sensing equipment).
Empty the syringe smoothly and completely until a soft click is heard. Repeat this cycle while observing the syringe strokes on the equipment screen. Do not bang the syringe when emptying it, to avoid any damage or errors. Tip: follow the screen or software instructions to aid performance of this procedure.

Linearity of the system is assessed as follows: The calibration manoeuver is typically performed three times at different flow rates. The recorded volume should be repeatable: when using a 3L syringe the largest - smallest measured volume should not deviate by more than 90 mL. Repeat above 3 to 5 times, you will be instructed to alter the speed of the air injections (slow, medium and fast) and must comply with the "multi flow" calibration standard (a range varying between 0.5 and 12 L/s with 3-L injection times between 0.5 and 6 s). The resulting simulated FVC values must fall within the +/- 3% range. Each time stamped successful calibration must be stored either within the software or as a print-out.
Calibration result The measured volume displacement should be within 2.91 L and 3.09 L at ATPS conditions. Results must be printed or stored for recall later.
Quality
Assurance Documentation
Device Quality Assurance

Attention to equipment quality assurance and calibration is an important part of good laboratory practice.

The minimum requirements are as follows:
Maintenance of a log of calibration results
Documentation of repairs or other alterations that return the equipment to acceptable operation
Recording of dates of computer software and hardware updates or changes
Recording the dates equipment is changed or relocated (e.g., industrial surveys)
Calibration verifications and quality control procedures must be repeated after any such changes before further testing begins

Table 1. Daily calibration procedure

 

Video 1. Spirometer calibration

 

Hygiene, Infection Protection and Control

  • The goal of infection control is to prevent the transmission of infection to patients and staff during pulmonary function testing.
  • Infection can be transmitted by direct contact with surfaces such as mouthpieces, noseclips, handheld spirometers, chair arms, and immediate proximal surfaces of valves or tubing.
  • Indirect transmission occurs by aerosol droplets generated by the patient blowing into the equipment but also expelled into the air of the testing room between manoeuvres.
Pre procedure notes Spirometry testing should be carried out in a suitable room and following local infection protection policy. It is advisable to have adequate fresh air ventilation and knowledge of the air changes per hour will inform the level of infection protection controls required. HVAC and portable HEPA air cleaners may be used to boost fresh air changes if needed.

Check patient symptoms for any possibility of transmission of infection. Postpone test if the patient is symptomatic e.g. influenza, RSV, COVID-19, current chest infection
Use of PPE is advised, and should be part of the pre-test risk assessment or patients and staff. The type and level of PPE used will be part of this assessment.

Hand hygiene is extremely important and must be performed by both the practitioner and patient before and after the test. Special attention to hand hygiene after touching surfaces and removal of PPE is also advised.

Educate the patient on cough etiquette and removal/replacement of their surgical facemask during the test (if required).
Check for any mobility issues or if they require assistance with using their facemask or staying on the mouthpiece during the test.
Consideration should be given to management of patients with special needs.

The testing equipment surfaces must be cleaned prior to use for every patient (Refer to manufacturer instruction for cleaning of equipment).
Single use disposable Bacterial /viral filters must be used for all tests including when using disposable sensors.

Follow flow chart below for more detail.
Patient factors Extra precautions should be taken for patients with, or suspected of having, tuberculosis, hemoptysis, oral lesions, or other known transmissible infectious diseases. Some patients are known to harbour virulent bacteria in their airways such as Pseudomonas aeruginosa or Pseudomonas sepacia. The equipment should be fully disinfected each time after use by such patients and not just the equipment surfaces. The same holds for patients with tuberculosis in whom bacteria are demonstrable in their sputum.

Immune-compromised patients and those with conditions such as HIV, cystic fibrosis or undergoing treatment cytostatic drugs, should be tested recently decontaminated equipment and prioritised for testing first of the day.

For high risk infected patients (who must be tested), possible precautions include;
Reserving equipment for the sole purpose of testing infected patients
Testing such patients at the end of the workday to allow time for spirometer disassembly and disinfection
Testing patients in their own rooms with adequate ventilation and appropriate protection for the operator.
Equipment factors Manufacturers must explicitly describe acceptable methods of cleaning and disinfecting their equipment, including recommended chemicals and concentrations, as well as safety precautions for the operator. Local infection control requirements, especially for at-risk populations such as patients with cystic fibrosis, may supersede both manufacturers' recommendations.
Perform procedure

Adapted from McGowan A et al. International consensus on lung function testing during the COVID-19 pandemic and beyond. ERJ Open Research 2022 8(1): 00602-2021; DOI: 10.1183/23120541.00602-2021
Post procedure During spirometry efforts the patient not only expels air, but also saliva sputum, oral debris. This contaminates the flow transducer of the spirometer and may cause it to malfunction. Use of single use filters will protect the sensors based on the bacterial and viral efficiency of the filter used. Most reusable flow transducers can be disassembled and should cleaned as per the manufacturers instructions. All disposable items, including filters, mouthpieces, noseclips, PPE and gloves, must be disposed of at the end of each patient testing session. Equipment and room surfaces must be cleaned between patients and comply with local policy.

Follow hand hygiene rules. (see Video 2 demonstration).

Table 2. Spirometry test hygiene, infection protection and control steps (should comply with local policy)

 

Video 2. Hand washing

 

Video 3. Special considerations

Special considerations include spirometry in the supine position, a child-friendly paediatric waiting room, the use of PPE with immunocompromised patients and proper disposal and disinfection (see Video 3 demonstration).